Also: Laser is shining in both arms (two light-sheets simultaneusly)
Possible reason: ThorLabs shutters are turned off or in manual mode
👉 You need to set each of the two ThorLabs shutter controllers to run on TTL input mode: after turning on the shutter controller, press and hold ENABLE butting for several seconds until the status LED blinks. Repeat this every time the shutter was turned off or the whole system was shut down, this is a common 'gotcha' point.
Also: you adjusted the histogram parameters to match the pixel values, but the image contains only noise
Possible reasons: Ludl filter wheel has its shutter closed; the filter position is incorrect; the assigned COM-port is incorrect and communication fails.
👉 Check if the Ludl filter wheel shutter is open (flip the shutter switches on its controller). Check if filter is actually moving when changed in mesoSPIM software. Check for COM-port errors in the mesoSPIM console. Compare COM ports in Windows Device manager with the port assigned to filter wheel in the mesoSPIM config file, and change the config file if necessary.
- 
Background
- mesoSPIM history
- Optical design
- Electronics
 
- 
Setting up a mesoSPIM
- First steps
- Preparing the software and electronics
- Preparing the microscope optics
- 
Setting the microscope up
- General alignment tips and tricks
- Installing the microscope base
- Setting up the detection path
- Alignment of the detection path
- Setup of the sample XYZ stages
- Setup of the excitation path
- Immersion cuvettes
- Set up a microscope config file
- Light-sheet co-alignment
- Set up initial ETL parameters
- Setting up lasers with the GUI
 
 
- Sample Handling
- Test Samples
- Troubleshooting
- Upgrades and custom variants