[NSRPARMA-131][feat] gui sparc: if laser interlock is triggered, also blank the beam #3009

Merged
Changes from all commits
Commits
File filter

Filter by extension

Filter by extension

Conversations
Failed to load comments.
Loading
Jump to
Jump to file
Failed to load files.
Loading
Diff view
Diff view
[feat] gui sparc: if laser interlock is triggered, also blank the beam
One of the most common reason for the laser interlock to be triggered
is that the user opened the SPARC ATM cover. The inside of the ATM can
receive X-rays caused by the e-beam. So, to be extra safe, also blank
the e-beam whenever the laser interlock is triggered.
  • Loading branch information
pieleric committed Mar 6, 2025
commit f718b81e16b881a13e57c0046210f3ea90473742
46 changes: 32 additions & 14 deletions src/odemis/gui/cont/acquisition/sparc_acq.py
Original file line number Diff line number Diff line change
Expand Up @@ -67,6 +67,12 @@ def __init__(self, tab_data, tab_panel, streambar_controller):
self._tab_panel = tab_panel
self._streambar_controller = streambar_controller
self._interlockTriggered = False # local/private bool to track interlock status
if self._main_data_model.ebeam and model.hasVA(self._main_data_model.ebeam, "blanker"):
# blanker state before interlock (to restore it when interlock is reset)
self._ebeam_blanker = self._main_data_model.ebeam.blanker
self._pre_interlock_blanker = self._ebeam_blanker.value
else:
self._ebeam_blanker = None

if model.hasVA(self._main_data_model.light, "interlockTriggered"):
# subscribe to the VA and initialize the warning status
Expand Down Expand Up @@ -217,33 +223,45 @@ def on_change_file(self, evt):
self.conf.fn_ptn, self.conf.fn_count = guess_pattern(new_name)
logging.debug("Generated filename pattern '%s'", self.conf.fn_ptn)

def on_interlock_change(self, value):
def on_interlock_change(self, value: bool) -> None:
"""
If the connected interlock status changes, the label on the acquisition panel has to
be updated. Either the status is not triggered and the notification label will not be
visible or the status is triggered and the user is notified through the label.
Update the label status through calling update_acquisition_time() without any arguments.
:param value (BooleanVA): current interlockTriggered VA value
:param value: current interlockTriggered VA value
"""
if value == self._interlockTriggered:
# if the value did not change in respect to the previous value
return

logging.warning(f"Interlock status changed from {not value} -> "
f"{value}")
logging.warning(f"Interlock status changed from {not value} -> {value}")

if value:
popup.show_message(wx.GetApp().main_frame,
title="Laser safety",
message=f"Laser was suspended automatically due to interlock trigger.",
timeout=10.0,
level=logging.WARNING)
# The laser is connected electronically to the interlock, so nothing to do in software.
# For the e-beam (to protect the user from X-rays), we force the blanker on.
if self._ebeam_blanker:
self._pre_interlock_blanker = self._ebeam_blanker.value
self._ebeam_blanker.value = True
message = "Laser and e-beam were suspended automatically due to interlock trigger."
else:
message = "Laser was suspended automatically due to interlock trigger."
else:
popup.show_message(wx.GetApp().main_frame,
title="Laser safety",
message=f"Laser interlock trigger is reset to normal.",
timeout=10.0,
level=logging.WARNING)
message = "Laser interlock trigger is reset to normal."
# Put back the e-beam blanker to its original state, but only if it is (still) forced
# on (otherwise, it's a sign that the user is manually handling it).
if self._ebeam_blanker and self._ebeam_blanker.value is True:
self._ebeam_blanker.value = self._pre_interlock_blanker
if self._pre_interlock_blanker is False:
message += " E-beam blanker disabled."
elif self._pre_interlock_blanker is None: # Automatic mode (= blanker active when not acquiring)
message += " E-beam blanker set back to automatic mode."

popup.show_message(wx.GetApp().main_frame,
title="Laser safety",
message=message,
timeout=10.0,
level=logging.WARNING)

self._interlockTriggered = value
self.update_acquisition_time()
Expand Down