Skip to content

ADALM2000 / m2k #2196

@Sharc99

Description

@Sharc99
  • W11
  • ADALM2000 Firmware v0.33
  • SW Version: SCOPY 2.1.0 rc2

Erratic/random behaviour on several PCs/Notebooks. Typical scenario:

  • Scopy crashes at 1st start
  • 2nd attempt: Scopy initiates up to Calibration, then crashes
  • Reset, reinstalling (to unprotected folders), deleting .ini and/or the ADI folder in the hidden AppData folder may or may not help. It behaves like a random pick
  • When SCOPY starts as expected (on rare occasions) the signal inputs seems to be disconnected (no Oscillator or Spectrum Analyzer input signals), while the outputs (generators) seem to be working
  • On rare occasions Scopy works as expected .... until the next start/connection
  • Weird ............

The former Scopy v 1.5.0 rc1 mostly works though.

Edit:
It seems like installing the drivers with v1.5.0 rc1 and then installing 2.1.0 rc2 skipping the drivers reinstall (or using the portable version) solved or mitigated some of the issues..... (?)