When a user enters carbs and delivers a bolus, a high temp immediately shows up as recommended. I believe this is because the meal is immediately taken into account by the Loop calculations, while the bolus is not, because the Loop relies solely on actual pump data to calculate IOB (as it should). This would not normally be an issue, as the Loop would become aware of the bolus delivered on the next pump readout. However, the high temp recommendation could become problematic if the user accepted the recommended high temp and if the Loop, for whatever reason, failed to communicate with the pump just after that. The user could be left with the bolus and the high temp for 30 min (effectively a double bolus) - a potentially hazardous situation.