Notifications
You must be signed in to change notification settings - Fork 1.4k
Description
Describe the bug
This morning I observed a mismatch between reported IOB and reported basal delivery:
Between 5:30 and 7am, Loop reports that it was delivering a neutral temp basal but that IOB rose from -0.5 to +1.0 over that time. This doesn't make sense... with a neutral temp (which would have been expected as my blood sugars were normal at the time), IOB should have been decaying towards zero.
Attach an Issue Report
Loop Report 2019-07-26 07:11:28-06:00.md.txt
Expected behavior
A clear and concise description of what you expected to happen.
Screenshots
See above
Phone
- Hardware: iPhone X
- OS Version: iOS 12.3
Loop Version
- Version Number: 1.10.0-dev, commit 5c63135
- Repo: LoopKit/Loop
CGM
- Device: Dexcom G6
- Manager app: Dexcom App
Pump
- Manufacturer: Insulet
- Model: Omnipod
- Firmware version: 2.8.0 / 2.8.0
Additional context
My basal rate this time of day is 1.35 U/hr, so I don't think this is a rounding / phase issue like #1015. It may be related to (or even the same as ?) #1034