Skip to content

EMU: CAEN V895 sending values on startup doesn't work consistently #5609

@Tom-Willemsen

Description

@Tom-Willemsen

As an instrument scientist on EMU, I would like the CAENV895 ioc to send the correct state on startup.

Currently it only sends the correct state for some channels (e.g. ch49 is ok, ch50 gets set to zero). I think this may be a race condition depending on the order of threshold and "enable" being sent. This results in no counts being recorded for some channels.

This is a requirement for EMU to be able to run IBEX in cycle

Acceptance criteria:

  • Restarting the CAEN V895 ioc on EMU puts the CAEN into the correct state.