Marker seems incorrect, FW unstable #40

Open
mare001 opened this issue Aug 26, 2020 · 1 comment
Open

Marker seems incorrect, FW unstable #40

mare001 opened this issue Aug 26, 2020 · 1 comment

Comments

@mare001
Copy link

mare001 commented Aug 26, 2020

It is not a big issue. Only if it helps for you:

If

  • more than one Marker are used (2, 3, or 4 Markers)
  • and stop frequency is higher as 2.14729 GHz
  • and the Marker is also placed higher as 2.14729 GHz
    the displayed value seem incorrect (with negative sign and wrong frequency)

In this setup (more than 1 marker and stop frequency higher than 2.14729 GHz) if stop frequency will changed one or more times, the firmware seems to be instable.

Used firmware: git-20200617-1a9a11d; Jun 19 2020 - 11:21:33
HW: ManoVNA V2_2

@mare001
Copy link
Author

mare001 commented Aug 26, 2020

Sorry, I saw:
This issue is discussed under "Fix frequency_string_short and its usage"

Thanks a lot for your software and NanoVNA V2!